In order to separate K and π with momenta up to 1.1 GeV/c using Time-of-Flight (TOF) over a distance of about 100 cm in the magnet, we will use Multi-Gap Resistive Plate Chamber (MRPC) due to the need of large area, high efficiency and overall time resolution in the range of 50-100 ps. We built a 260 μm × 10 gaps MRPC with an active area of 10 × 50 cm2 and a pad size of 2.5 × 40 cm2. We studied the trigger rate dependence and the gas mixture of this MRPC. It was found that high enough voltage is necessary to achieve a good time resolution and high efficiency at large trigger rates. The gas mixture composed of R134a : SF6 : Iso-Butane at 90% : 5% : 5% is our optimized choice.
原文 | ???core.languages.en_GB??? |
---|---|
文章編號 | 061 |
期刊 | Proceedings of Science |
卷 | 2012-February |
出版狀態 | 已出版 - 2012 |
事件 | 11th Workshop on Resistive Plate Chambers and Related Detectors, RPC 2012 - Rome, Italy 持續時間: 5 2月 2012 → 10 2月 2012 |