Test results of the L3 precision muon detector

B. Adeva, M. Aguilar-Benitez, A. Aloisio, M. G. Alviggi, D. Antreasyan, L. Baksay, U. Becker, J. Berdugo, G. J. Bobbink, J. G. Branson, C. Burgos, J. D. Burger, D. Campana, F. Carbonara, M. Cerrada, Y. H. Chang, M. L. Chen, G. Chiefari, N. Colino, E. DragoP. Duinker, I. Duran, F. Erné, R. Fabbretti, M. Fabre, S. Feher, S. S. Gau, M. Gettner, E. Gonzalez, H. van der Graaf, D. Güsewell, F. G. Hartjes, V. Innocente, B. N. Jin, A. König, S. Lanzano, J. M. LeGoff, X. Leijtens, L. Li, S. Lokos, C. Mana, L. Martinez, G. G.G. Massaro, Y. F. Mao, R. Magahiz, P. McBride, L. Merola, M. Napolitano, H. Onvlee, G. Paternoster, S. Patricelli, Y. Peng, O. Prokofiev, J. M. Qian, P. Razis, L. Romero, J. A. Rubio, H. Rykaczewski, J. Salicio, K. Schetkovsky, P. Schmitt, C. Sciacca, P. G. Seiler, V. Souvorov, K. Strauch, C. Timmermans, S. Volkov, J. H. Wang, Q. F. Wang, Z. M. Wang, T. Wenaus, M. White, R. Wilhelm, C. Willmott, B. Wyslouch, F. Y. Xi, B. Z. Yang, C. G. Yang, Q. Y. Yang, C. H. Ye, S. C. Yeh, C. R. Zabounidis, Z. P. Zhang

研究成果: 雜誌貢獻期刊論文同行評審

5 引文 斯高帕斯(Scopus)

摘要

The L3 detector is designed to measure the muon momentum with a 2% resolution at p = 45 GeV/c. We discuss here the systems we developed to reach the required accuracy and control the mechanical alignment at running time. We also report on the test done on the muon spectrometer with UV lasers and cosmic rays.

原文???core.languages.en_GB???
頁(從 - 到)335-341
頁數7
期刊Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
289
發行號3
DOIs
出版狀態已出版 - 15 4月 1990