@inproceedings{43921481b4f544368092c9510cb9329d,
title = "Rotational scanning mechanism of multi-point synchronous detection for near infrared diffuse optical tomography",
abstract = "This study aims at the design, implementation, and verification of a near infrared (NIR) diffuse optical tomography (DOT) electro-optical imaging system based on a measuring instrument with scanning mechanism of synchronous multi-point detection and an image reconstruction scheme implemented on an off-line PC. The developed measurement instrument possesses a high degree of angular resolution and can be operated in a continuous-intensity or intensity modulation scheme to acquire output NIR radiance signals. Experimental trials are conducted using both homogeneous and heterogeneous phantoms made of high-scattering Intralipid. Promising results have been obtained with a superior spatial resolution, especially for absorption image, compared with the system with fixed number of detection channels.",
author = "Pan, {Min Chun} and Chen, {Chien Hung} and Chen, {Lian Yu} and Pan, {Min Cheng}",
year = "2008",
doi = "10.1115/biomed2008-38073",
language = "???core.languages.en_GB???",
isbn = "0791848337",
series = "2008 Proceedings of the 3rd Frontiers in Biomedical Devices Conference and Exhibition",
publisher = "ASME",
pages = "107--108",
booktitle = "2008 Proceedings of the 3rd Frontiers in Biomedical Devices Conference and Exhibition",
note = "2008 3rd Frontiers in Biomedical Devices Conference and Exhibition ; Conference date: 18-06-2008 Through 20-06-2008",
}