This study investigated the flow effect on the lesion formation during radio-frequency cardiac catheter ablation in temperature-controlled mode. The blood flow in heart chambers carries heat away from the endocardium by convection. This cooling effect requires more power from the ablation generator and causes a larger lesion. We set up a flow system to simulate the flow inside the heart chamber. We performed in vitro ablation on bovine myocardium with three different flow rates (0 L/min, 1 L/rain and 3 L/min) and two target temperatures (60 °C and 80 °C). During ablation, we also recorded the temperatures inside the myocardium with a three-thermocouple temperature probe. The results show that lesion dimensions (maximum depth, maximum width and lesion volume) are larger in high flow rates (p < 0.01). Also, the temperature recordings show that the tissue temperature rises faster and reaches a higher temperature under higher flow rate.
原文 | ???core.languages.en_GB??? |
---|---|
頁(從 - 到) | 425-433 |
頁數 | 9 |
期刊 | IEEE Transactions on Biomedical Engineering |
卷 | 48 |
發行號 | 4 |
DOIs | |
出版狀態 | 已出版 - 2001 |