摘要
Aspartate aminotransaminase (AST) is a hepatocelluar enzyme released into the bloodstream when hepatic cells are damaged, resulting in elevated blood levels of AST. A single use, disposable biosensor prototype, composed of catalytic iridium nano-particles dispersed on carbon paste, was developed to detect enzymatically-produced H2O2 in AST-mediated reactions. This biosensor is capable of measuring AST levels in a phosphate buffer and undiluted human serum over the concentration range of 0 to 0.89 μg/mL AST concentration (corresponding to 0-250 UL-1 specific activity). The biosensor operates at relatively low oxidation potential (+0.3 volt (V) versus the printed Ag/AgCl), minimizing any potential chemical interference in human serum. The measurements of AST in human serum using the biosensor compared well with those measured by standard hospital spectrophotometric assays. This Ir-C biosensor may be useful for AST measurements in the clinical environment.
原文 | ???core.languages.en_GB??? |
---|---|
頁(從 - 到) | 234-244 |
頁數 | 11 |
期刊 | Biosensors |
卷 | 2 |
發行號 | 2 |
DOIs | |
出版狀態 | 已出版 - 6月 2012 |