摘要
As in the lens optimization process, ophthalmic lens designers have usually constructed error functions with 0.5, 0.7, and 1.0 at only three oblique fields. This seems enough to achieve a balanced trade-off with the astigmatic error, the power error, and the distortion all being considered simultaneously. However, for high-power ophthalmic lenses, the aberration curves show serious violations even if aspherical coefficients are involved. The analytical results indicate that a field error suppression of up to 7 points may even be required in some cases. The suppression effects are excellent and examples of both positive and negative lenses are designed.
原文 | ???core.languages.en_GB??? |
---|---|
頁(從 - 到) | 460-470 |
頁數 | 11 |
期刊 | Optical Engineering |
卷 | 41 |
發行號 | 2 |
DOIs | |
出版狀態 | 已出版 - 2月 2002 |