摘要
Microfabricated platinum redox sensors can be used for indication in redox diffusional microtitration of microliter size body fluid samples. The sensor can be deposited onto different substrate materials. Adsorption of reagent and/or sample on alumina ceramic substrate versus Pyrex glass substrate is compared. It is demonstrated that in a classical iodimetric redox titration system, significant adsorption occurs on alumina ceramic substrate, causing irreproducible results. Reproducibility with a 1.7% coefficient of variation (CV) is achieved using a sensor deposited on Pyrex substrate for the analysis of 20 μL sample droplets.
原文 | ???core.languages.en_GB??? |
---|---|
頁(從 - 到) | 824-825 |
頁數 | 2 |
期刊 | Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings |
卷 | 16 |
發行號 | pt 2 |
出版狀態 | 已出版 - 1994 |
事件 | Proceedings of the 16th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. Part 1 (of 2) - Baltimore, MD, USA 持續時間: 3 11月 1994 → 6 11月 1994 |