@inproceedings{7a8505dbb4fb482694387454c0815549,
title = "Beam profile monitoring system for proton therapy",
abstract = "An innovative design of 16-channel front-end electronics for ionization chamber in monitoring proton beam profile has been developed to feature simplicity, low cost, durability, and reliability without compromising its performances. All the parts are available off the shelf. The front-end electronics board is powered by USB connection, through which the data are transferred to the host computer for storage and further online analysis. The system has been studied with Geant4 MC [1] and tested with both electron and proton beams.",
author = "Chili Ho and Chen, {Augustine E.} and Shujhen Dai and Minglee Chu and Lin, {Chih Hsun} and Teng, {Ping Kun} and Wang, {Chung Hsiang} and Chao, {Tsi Chian} and Lee, {Chung Chi} and Tung, {Chuang Jong} and Duh, {Ting Shien}",
year = "2010",
doi = "10.1109/NSSMIC.2010.5873797",
language = "???core.languages.en_GB???",
isbn = "9781424491063",
series = "IEEE Nuclear Science Symposium Conference Record",
pages = "440--442",
booktitle = "IEEE Nuclear Science Symposuim and Medical Imaging Conference, NSS/MIC 2010",
note = "2010 IEEE Nuclear Science Symposium, Medical Imaging Conference, NSS/MIC 2010 and 17th International Workshop on Room-Temperature Semiconductor X-ray and Gamma-ray Detectors, RTSD 2010 ; Conference date: 30-10-2010 Through 06-11-2010",
}