Abstract
An automated gas chromatographic system was constructed to easily adapt either the cryogenic trap or chemical sorbent trap for preconcentrating ambient levels of volatile organic compounds. Remarkable similarity in chromatograms from C3 to C10 was found between these two enrichment methods, except that the sorbent trap did not quantitatively trap the C2-hydrocarbons. In contrast to cryogenic trapping, the chromatographic conditions for more volatile compounds were substantially improved using the sorbent trap. Water interference on the porous-layer open tubular column was also better managed using the sorbent trap for the continuous analysis of humid room air. The similarity in peak profiles between the GC-flame ionization detection (FID) and a commercial GC-MS system, regardless of concentration levels, facilitated compound identification on the FID chromatograms based on a field mission involving analysis of 106 air samples. Copyright (C) 1999 Elsevier Science B.V.
Original language | English |
---|---|
Pages (from-to) | 183-193 |
Number of pages | 11 |
Journal | Journal of Chromatography A |
Volume | 863 |
Issue number | 2 |
DOIs | |
State | Published - 26 Nov 1999 |
Keywords
- Air analysis
- Environmental analysis
- Volatile organic compounds