Abstract
We outline the design of a new small-animal PET system. This system employs modular scintillation cameras composed of nine PMTs coupled by a light guide to a monolithic NaI(Tl) scintillation crystal. The basic principles of the PET system are presented along with an overview of the electronics and position-estimation scheme. The potential advantages of this design are discussed and preliminary reconstructed images are presented.
Original language | English |
---|---|
Article number | M9-47 |
Pages (from-to) | 3389-3392 |
Number of pages | 4 |
Journal | IEEE Nuclear Science Symposium Conference Record |
Volume | 6 |
State | Published - 2004 |